Title: CT-MRI automatic surface-based registration schemes combining global and local optimization techniques
Authors: Ματσόπουλος, Γεώργιος Κ.
Δελήμπασης, Κωνσταντίνος Κ.
Μουραβλιάνσκυ, Νικόλαος Α.
Ασβεστάς, Παντελής Α.
Νικήτα, Κωνσταντίνα Σ.
Contributors: Κουλουλιάς, Βασίλειος Ε.
Ουζούνογλου, Νικόλαος Κ.
Item type: Journal article
Keywords: Medical image registration;Clinical environment;Ιατρική εικόνα;Κλινικό περιβάλλον
Subjects: Medicine
Medical technology
Ιατρική
Ιατρικά όργανα και εξοπλισμός
Issue Date: 13-Jan-2015
2003
Publisher: IOS Press
Abstract: Medical image registration is commonly required in order to combine the complementary information provided by different medical imaging modalities. In this paper, a new automatic registration scheme is proposed to register 3-D CT-MR head images and is currently tested on a clinical environment. The proposed scheme, after the preprocessing and the outer surface extraction of the data, is based on the use the rigid transformation method, in conjunction with a combination of global and local optimization techniques. Analytically, the paper exploits the optimization efficiency of three widely used optimization techniques, in obtaining the parameters of the rigid transformation model: the Downhill Simplex Method, the Genetic Algorithms and the Simulated Annealing. These optimization techniques are further combined by the sequential application of the Powell optimization method in order to refine the registration and increase its accuracy. A comparative study involving these optimization techniques in conjunction with the rigid transformation, and two other methods, the ICP and the manual methods, is also presented, for a sufficient number of clinical CT-MR brain images. Finally, quantitative and qualitative results are also presented to validate the performance of these automatic surface-based registration schemes, in terms of consistency and accuracy. Throughout of this study, the automatic registration scheme comprising of the rigid transformation in conjunction with the Simulated Annealing method sequentially combined with the Powell method has been performed superior regarding all the other compared registration schemes.
Language: English
Citation: Matsopoulos, K., Delibasis, K., Mouravliansky, N., Asvestas, P., Nikita, K., et al. (2003). CT-MRI automatic surface-based registration schemes combining global and local optimization techniques. Technology and Health Care. 11(4). pp. 219-232. IOS Press.
Journal: Technology and Health Care
Type of Journal: With a review process (peer review)
Access scheme: Embargo
License: Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες
URI: http://hdl.handle.net/11400/3939
Appears in Collections:Δημοσιεύσεις

Files in This Item:
File Description SizeFormat 
CT-MRI automatic surface-based registration.pdf
  Restricted Access
177.04 kBAdobe PDFView/Open Request a copy


This item is licensed under a Creative Commons License Creative Commons