Title: Comparison of full field digital (FFD) and computed radiography (CR) mammography systems in Greece
Authors: Καλαθάκη, Μαρία
Χουρδάκης, Κωνσταντίνος
Οικονομίδης, Στέφανος
Τριτάκης, Π.
Καλύβας, Νεκτάριος Ι.
Contributors: Σημαντηράκης, Γεώργιος
Μανουσαρίδης, Γ.
Καϊσάς, Ι.
Καμενοπούλου, Βασιλική
Item type: Journal article
Keywords: Full field digital;Radiography, Medical--Digital techniques;Πλήρες ψηφιακός τομέας;Υπολογιστική ακτινογραφία
Subjects: Medicine
Biomedical engineering
Ιατρική
Βιοϊατρική τεχνολογία
Issue Date: 20-Jan-2015
2011
Date of availability: 20-Jan-2015
Publisher: Oxford University Press
Abstract: The purpose of this study is to evaluate and compare the performance of 52 full field digital (FFD) and computed radiography (CR) mammography systems checked by the Greek Atomic Energy Commission with respect to dose and image quality. Entrance surface air kerma (ESAK) was measured and average glandular dose (AGD) was calculated according to the European protocol on dosimetry in mammography. The exposures were performed using the clinical protocol of each laboratory. The image quality was assessed by the total score of resolved phantom structures incorporated in an American College of Radiology accreditation phantom. The mean ESAK values for FFD and CR systems were 4.59±1.93 and 5.0±1.78 mGy, respectively, whereas the AGD yielded a mean value of 1.06±0.36 mGy for the FFD and 1.04±0.35 mGy for the CR systems. Considering image quality, FFD systems indicated a mean total score of 13.04±0.89, whereas CR systems a mean total score of 11.54±1.06.
Language: English
Citation: Kalathaki, M., Hourdakis, C., Economides, S., Tritakis, P., Kalyvas, N., et al. (August 2011). Comparison of full field digital (FFD) and computed radiography (CR) mammography systems in Greece. Radiation Protection Dosimetry. 147(1-2). pp. 202-205. Oxford University Press. Available from: http://rpd.oxfordjournals.org [Accessed 04/08/2011]
Journal: Radiation Protection Dosimetry
Type of Journal: With a review process (peer review)
Access scheme: Publicly accessible
License: Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες
URI: http://hdl.handle.net/11400/4325
Appears in Collections:Δημοσιεύσεις

Files in This Item:
File Description SizeFormat 
RPD_mammo.pdf118.69 kBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons