Title: | Experimental determination of luminescence emission properties of lyso: CE and CSI: TL single crystal scintillators under uv and x-ray exposures for use in medical imaging |
Authors: | Βαλαής, Ιωάννης Γ. Κανδαράκης, Ιωάννης Σ. Νικολόπουλος, Δημήτριος Λούντος, Γεώργιος Γιόκαρης, Νικόλαος |
Contributors: | Καραγιάννης, Κωνσταντίνος Επισκοπάκης, Αναστάσιος Δημητρόπουλος, Νικόλαος Παναγιωτάκης, Γεώργιος Σ. |
Item type: | Conference publication |
Conference Item Type: | Full Paper |
Keywords: | Inorganic scintillators;Nuclear counters;Ανόργανοι σπινθηριστές;Ανιχνευτές ακτινοβολίας |
Subjects: | Medicine Physics Ιατρική Φυσική |
Issue Date: | 22-Jan-2015 2005 |
Publisher: | [χ.ό.] |
Abstract: | Cesium iodide doped with thallium, CsI: Tl, is a high efficiency scintillator yielding high resolution images. It has been used in several diagnostic equipment, especially in fluoroscopy, in single photon emission computed tomography (SPECT) and in x-ray computed tomography (CT) systems. Lutetium-yttrium based scintillators, such as LYSO: Ce, are high effective atomic number, non-hygroscopic, fast emitting materials, promising candidates for use in positron emission (PET) imagers. The present study investigates the light emission characteristics of LYSO: Ce and CsI: Tl single crystal scintillators under UV and x-ray medical imaging conditions. To this aim parameters related to the luminescence emission spectrum and emission efficiency were studied using experimental methods. Various x-ray tube voltages currently employed in x-ray imaging techniques were used. In addition the emission spectrum under UV and x-ray excitation was used to determine the spectral compatibility to optical photon detectors incorporated in medical imaging systems. LYSO: Ce was found of equally high efficiency, however LYSO: Ce may be considered more suitable for computerized tomographic applications due to its very fast temporal response. |
Language: | English |
Citation: | Valais, I., Kandarakis, I., Nikolopoulos, D., Loudos, G., Giokaris, N., et al. (2005). Experimental determination of luminescence emission properties of lyso: CE and CSI: TL single crystal scintillators under uv and x-ray exposures for use in medical imaging. In the 1st International Conference on Experiments/Process/System Modelling/Simulation/Optimization. Athens, 6th-9th July 2005. |
Conference: | International Conference on Experiments/Process/System Modelling/Simulation/Optimization |
Access scheme: | Embargo |
License: | Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες |
URI: | http://hdl.handle.net/11400/4542 |
Appears in Collections: | Δημοσιεύσεις |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
epsmso05_110.pdf Restricted Access | 19.58 kB | Adobe PDF | View/Open Request a copy |
This item is licensed under a Creative Commons License