Title: A prototype device for standardized calibration of pulse oximeters
Authors: Hornberger, C.
Knoop, Ph.
Nahm, W.
Matz, H.
Konecny, E.
Contributors: Μπενέκος, Κ.
Βαλαής, Ιωάννης Γ.
Avgerinos, J.
Καρούτης, Α.
Ικιάδης, Α.
Gehrin, H.
Bonk, R.
Frankenberger, H.
Meyfroidt, G.
Wouters, P.
Gil-Rodriguez, J.
Ponz, L.
Weininger, S.
Item type: Journal article
Keywords: Pulse oximetry;Infrared equipment--Calibration--Congresses;Παλμική οξυμετρία;Βαθμονόμηση
Subjects: Medicine
Medical technology
Ιατρική
Ιατρική τεχνολογία
Issue Date: 23-Jan-2015
2000
Date of availability: 23-Jan-2015
Publisher: Kluwer Academic Publishers
Abstract: Objective.To develop and test a method for standardized calibrationof pulse oximeters. Methods.A novel pulse oximeter calibrationtechnique capable of simulating the behavior of real patients is discussed.It is based on an artificial finger with a variable spectral-resolved lightattenuator in conjunction with an extensive clinical database of time-resolvedoptical transmission spectra of patients fingers in the wavelength range600–1000 nm. The arterial oxygen saturation of the patients at the timeof recording was derived by analyzing a corresponding blood sample with aCO-oximeter. These spectra are used to compute the modulation of the lightattenuator which is attached to the artificial finger. This calibration methodwas tested by arbitrarily playing back recorded spectra to pulse oximeters andcomparing their display to the value they displayed when the spectra wererecorded. Results.We were able to demonstrate that the calibratorcould generate physiological signals which are accepted by a pulse oximeter.We also present some experience of playing back recorded patient spectra. Themean difference between the original reading of the pulse oximeters and thedisplay when attached to the calibrator is 1.2 saturation points (displayedoxygen saturation SpO2) with a standard deviation of 1.9 saturationpoints. Conclusions.The tests have shown the capabilities of aspectral light modulator for use as a possible calibration standard for pulseoximeters. If some improvements of the current prototype can be achieved weconclude from the experience with the device that this novel concept for thecalibration of pulse oximeters is feasible and that it could become animportant tool for assessing the performance of pulse oximeters.
Language: English
Citation: Hornberger, C., Knoop, P., Nahm, W., Matz, H., Konecny, E., et al. (March 2000). A prototype device for standardized calibration of pulse oximeters. Journal of Clinical Monitoring and Computing. 16(3). pp. 161-169. Kluwer Academic Publishers. Available from: http://link.springer.com [Accessed 0&/2006]
Journal: Journal of Clinical Monitoring and Computing
Type of Journal: With a review process (peer review)
Access scheme: Publicly accessible
License: Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες
URI: http://hdl.handle.net/11400/4596
Appears in Collections:Δημοσιεύσεις

Files in This Item:
File Description SizeFormat 
1-POC_JCMC_2000.pdf191.94 kBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons