Please use this identifier to cite or link to this item: http://hdl.handle.net/11400/5594
Title: Noise monitoring and alarm system for infant incubators
Authors: Στρατικόπουλος, Ηλίας Σ.
Σαατσάκης, Γεώργιος
Καλλέργη, Μαρία
Item type: Conference publication
Keywords: Incubator noise;Neurological disorders;Θόρυβος Θερμοκοιτίδας;Νευρολογικές διαταραχές;Υγεία του βρέφους;Infant health;Digital detector;Automated control;Αυτόματος έλεγχος;Ψηφιακός ανιχνευτής
Subjects: Medicine
Pediatrics
Ιατρική
Παιδιατρική
Issue Date: 3-Feb-2015
2012
Date of availability: 3-Feb-2015
Publisher: Νερατζής, Ηλίας
Σιανούδης, Ιωάννης
Βαλαής, Ιωάννης Γ.
Φούντος, Γεώργιος Π.
Abstract: Clinical studies have demonstrated a correlation between high noise levels in a neonatal intensive care unit (NICU) and health disorders in newborns. This work aimed at: (a) performing a qualitative and quantitative analysis of noise in the interior and exterior of NICU incubators, and (b) developing and evaluating a system to monitor noise and alert the NICU staff. The noise levels inside and outside the incubators of an NICU were recorded for three consecutive days. An automatic control system was also developed and clinically tested to monitor and record noise and generate a visual alarm when levels were above a selected threshold. Noise levels were above recommended standards inside and outside the incubators. The external noise sources were the routine activities of NICU personnel and the operation of other medical systems in the unit. The internal noise sources were the temperature and ventilation systems of the incubators. The proposed noise monitoring and alarm system could improve infant care. The system could raise awareness and improve NICU personnel training thus, effectively addressing one major, external source of noise. The system could also provide valuable data to the manufacturers, who, in the long run, could improve the technical specifications of the incubators.
Description: Special issue: Workshop on Bio-Medical Instrumentation and related Engineering And Physical Sciences, Technological Educational Institute of Athens, 6 July 2012
Language: English
Citation: Stratikopoulos, I.S., Saatsakis, G. and Kallergi, M. (2012). Noise monitoring and alarm system for infant incubators. "e-Journal of Science & Technology". [Online] 7(3): 73-77. Available from: http://e-jst.teiath.gr/
Journal: e-Περιοδικό Επιστήμης & Τεχνολογίας
e-Journal of Science & Technology
Type of Journal: With a review process (peer review)
Access scheme: Publicly accessible
License: Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες
URI: http://hdl.handle.net/11400/5594
Appears in Collections:Τόμος 07, τεύχος 3 (2012)

Files in This Item:
File Description SizeFormat 
Stratikopoulos.pdf252.36 kBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons